Health Canada · November 13, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20
- Category
- Medical device recalls
- Hazard
- Philips has identified two software issues affecting Azurion R3.1 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement. Issue 1 - System remains in continuous restart mode after the start-up: Due to a software…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20 identifiers →Health Canada · November 13, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20
- Category
- Health product
- Hazard
- Philips has identified two software issues affecting Azurion R3.1 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement. Issue 1 - System remains in continuous restart mode after the start-up: Due to a software…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20 identifiers →Health Canada · October 29, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Azurion System
- Category
- Medical device recalls
- Hazard
- Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Azurion System identifiers →Health Canada · October 29, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Azurion System
- Category
- Health product
- Hazard
- Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Azurion System identifiers →Health Canada · October 21, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 Systems
- Category
- Health product
- Hazard
- Philips Azurion systems use solid-state drives (SSDs) in various PCs to store the operating system and image data and to support system functionality. Philips has identified that, during certain repair or service actions, SSDs that are not specified in the…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 Systems identifiers →Health Canada · October 21, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 Systems
- Category
- Medical device recalls
- Hazard
- Philips Azurion systems use solid-state drives (SSDs) in various PCs to store the operating system and image data and to support system functionality. Philips has identified that, during certain repair or service actions, SSDs that are not specified in the…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 Systems identifiers →Health Canada · September 17, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Ingenia, Achieva and Intera MR
- Category
- Product recall
- Hazard
- When exporting diffusion tensor imaging (DTI) scans from a Philips MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Ingenia, Achieva and Intera MR identifiers →Health Canada · September 9, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Veradius Unity and BV Pulsera Mobile C-Arm
- Category
- Medical device recalls
- Hazard
- Air bubbles may develop in the monoblock oil when there is a prolonged period between manufacture and installation, for example due to storage, transportation, or other pre-installation holding time. During a clinical procedure, when the user activates X-ray…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Veradius Unity and BV Pulsera Mobile C-Arm identifiers →Health Canada · September 9, 2026
- Affected product
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Ingenia, Achieva and Intera MR
- Category
- Medical device recalls
- Hazard
- Philips has become aware of a software issue affecting Philips Magnetic Resonance (MR) imaging systems operating with software versions R5 & R11, which may terminate an examination during the image reconstruction process. If the scan parameters are manually…
Check affected PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Ingenia, Achieva and Intera MR identifiers →