PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20
OFFICIAL-SOURCE RECALL · CANADA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20 Recall: Philips has identified two software issues affecting
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Azurion 7 M20 recall in Canada: Philips has identified two software issues affecting Azurion R3.1. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Azurion 7 M20
- Manufacturer
- See official notice
- Model number
- 722234
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Philips has identified two software issues affecting Azurion R3.1 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement. Issue 1 - System remains in continuous restart mode after the start-up: Due to a software issue, the License Manager - which verifies installed licenses and enables optional functionality in the Azurion system based on the available licenses - expands the Windows registry database at every system start or restart. Over time, the registry may grow beyond the maximum size that the License Manager can load. When this happens, the License Manager cannot start and the system will enter a continuous reboot loop. The Azurion system will not complete booting, and the message "X-ray system is booting..." remains displayed on the screen. Issue 2 - Longitudinal position error - Applicable only to azurion systems with Poly-G3 frontal stand: When requesting a movement of the frontal stand in longitudinal direction, the Azurion system may detect a mismatch between the expected longitudinal (set-) position and the measured (actual) position of the stand. This mismatch is caused by a glitch in the longitudinal position value provided by the position sensing potentiometer. When the mismatch is detected, the motion power to the frontal stand is automatically cut off. The message "Some stand movements are not available" is displayed to the user. In this situation, X-ray imaging and table movements will continue to function. This issue can be resolved with a cold restart of the system or by restarting the geometry.
Hazard
Philips has identified two software issues affecting Azurion R3.1 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement. Issue 1 - System remains in continuous restart mode after the start-up: Due to a software issue, the License Manager - which verifies installed licenses and enables optional functionality in the Azurion system based on the available licenses - expands the Windows registry database at every system start or restart. Over time, the registry may grow beyond the maximum size that the License Manager can load. When this happens, the License Manager cannot start and the system will enter a continuous reboot loop. The Azurion system will not complete booting, and the message "X-ray system is booting..." remains displayed on the screen. Issue 2 - Longitudinal position error - Applicable only to azurion systems with Poly-G3 frontal stand: When requesting a movement of the frontal stand in longitudinal direction, the Azurion system may detect a mismatch between the expected longitudinal (set-) position and the measured (actual) position of the stand. This mismatch is caused by a glitch in the longitudinal position value provided by the position sensing potentiometer. When the mismatch is detected, the motion power to the frontal stand is automatically cut off. The message "Some stand movements are not available" is displayed to the user. In this situation, X-ray imaging and table movements will continue to function. This issue can be resolved with a cold restart of the system or by restarting the geometry.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Azurion 7 M20
Agency: Health Canada
Official publication date: November 13, 2026