PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Azurion System Recall: Philips has identified a safety issue affecting
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Philips Azurion System recall in Canada: Philips has identified a safety issue affecting certain. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Philips Azurion System
- Manufacturer
- See official notice
- Model number
- 722 224, 722 079, 722 067, 722 223, 722 078, 722 226, 722 068
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system. When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use.
Hazard
Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system. When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use.
What consumers should do
Contact the manufacturer if you require additional information.