Therapeutic Goods Administration · August 7, 2026
- Affected product
- Pertuzumab 420mg in 0.9% Sodium Chloride in Polyolefin Bag (Freeflex) for
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Cairns Hospital reported on 11/08/2026 that a Pertuzumab infusion bag (BB26084718) had expired prior to treatment. The single dose unit had expired during transit to the customer. The customer advised the single…
Check affected Pertuzumab 420mg in 0.9% Sodium Chloride in Polyolefin Bag (Freeflex) for identifiers →Therapeutic Goods Administration · August 7, 2026
- Affected product
- StrykeFlow PS and AHTO Tube Sets
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action There has been an increase in reported complaints describing leaking from the device during use. During use, the device may experience fluid leakage from the handpiece, which may result in fluid accumulating on the…
Check affected StrykeFlow PS and AHTO Tube Sets identifiers →Therapeutic Goods Administration · August 6, 2026
- Affected product
- Bravo CF capsule delivery device, 5-pk
- Category
- Medical device recalls
- Hazard
- Classification Class I Reason for the Action Medtronic has identified an increase in complaints involving the Bravo CF capsule delivery device where the capsule fails to attach to the oesophagus or detach from the delivery device. Medtronic's preliminary root…
Check affected Bravo CF capsule delivery device, 5-pk identifiers →Therapeutic Goods Administration · August 6, 2026
- Affected product
- Meisinger Polishing 2 Step DPO02 204 040 / 2
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Surface cracks of varying severity may occur in the affected polishers. Depending on their extent, these cracks may be visible to different degrees and may impair the mechanical stability of the instrument.…
Check affected Meisinger Polishing 2 Step DPO02 204 040 / 2 identifiers →Therapeutic Goods Administration · August 5, 2026
- Affected product
- Craft Botanics White Runtz T25 10g Jar (Dried Herb)
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action A labelling error has been identified for a specific batch of Craft Botanics White Runtz, following a customer complaint about a different organoleptic profile (appearance and aroma) compared to previous batches.…
Check affected Craft Botanics White Runtz T25 10g Jar (Dried Herb) identifiers →Therapeutic Goods Administration · August 5, 2026
- Affected product
- WaveWriter Alpha Spinal Cord Stimulation (SCS) System
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action Boston Scientific is notifying customers of an update to the Information for Prescribers and Information for Patients manuals used in Australia for applicable Boston Scientific WaveWriter Alpha Spinal Cord…
Check affected WaveWriter Alpha Spinal Cord Stimulation (SCS) System identifiers →Therapeutic Goods Administration · August 4, 2026
- Affected product
- Percutaneous Insertion Kits (PIK)
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action It was identified that there have been failures in sterile barrier integrity verification testing. As these failures were not previously addressed in accordance with packaging verification requirements, the…
Check affected Percutaneous Insertion Kits (PIK) identifiers →Therapeutic Goods Administration · August 4, 2026
- Affected product
- Fludeoxyglucose (18F) Injection (FDG)
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action It was identified there was a discrepancy of the recorded activity and volume supplied on the QC Release Forms than the actual activity and volume supplied in the customer vials. As such, the vials may not have met…
Check affected Fludeoxyglucose (18F) Injection (FDG) identifiers →Therapeutic Goods Administration · August 3, 2026
- Affected product
- GS1-HT Extraction Script for use within EasyScreen Sample Processing Kit Workflow.
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action The impacted extraction scripts contain a missing section of error-handling code for sample positions 81–95. If a low-volume or missing-sample condition occurs in positions 81 - 95, the script may apply recovery…
Check affected GS1-HT Extraction Script for use within EasyScreen Sample Processing Kit Workflow. identifiers →Therapeutic Goods Administration · August 3, 2026
- Affected product
- OPTIRAY 350 Ioversol 74% w/v 74.1g/100mL Injection Vial
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action The vial label of some batches of Optiray 350 100 mL vials is incorrect. It lists the ingredients' quantities for Optiray 350 75 mL vial instead. The ingredients' quantities listed on the product carton are…
Check affected OPTIRAY 350 Ioversol 74% w/v 74.1g/100mL Injection Vial identifiers →Therapeutic Goods Administration · August 3, 2026
- Affected product
- XScribe Cardiac Stress Testing System
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action Some Q-Stress and XScribe systems experience a problem with the Windows 11 image installed on affected computers. Users may observe a Windows system warning message upon startup that prevents the installation of…
Check affected XScribe Cardiac Stress Testing System identifiers →Therapeutic Goods Administration · August 3, 2026
- Affected product
- LATITUDE Clarity Server Software
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action There is a potential for a delay in AF Burden Alert notification affecting the LATITUDE Clarity server software which supports the LUX-Dx family of ICMs. This delayed AF Burden Alert notification behaviour by…
Check affected LATITUDE Clarity Server Software identifiers →Therapeutic Goods Administration · July 31, 2026
- Affected product
- GYNECARE TVT EXACT Continence System
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action The manufacturer, Ethicon SARL, has identified a problem whereby white particulates were observed within the sterile barrier of certain LOTs. The manufacturer has determined that the white particulates originated…
Check affected GYNECARE TVT EXACT Continence System identifiers →Therapeutic Goods Administration · July 31, 2026
- Affected product
- Mindray Monitors
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Mindray has identified that certain monitor manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions. To improve clarity and ensure…
Check affected Mindray Monitors identifiers →Therapeutic Goods Administration · July 30, 2026
- Affected product
- Beckman Coulter Chemistry Analyzer AU480. An in vitro diagnostic medical device
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Beckman Coulter has determined that CHEMISTRY ANALYZER AU480 with Windows 10 IoT Enterprise 2021 LTSC PCs listed above may output the message “System Stopped By Fatal Error: DBConnectionError AlarmDB when launching…
Check affected Beckman Coulter Chemistry Analyzer AU480. An in vitro diagnostic medical device identifiers →Therapeutic Goods Administration · July 29, 2026
- Affected product
- High Filtration Surgical Face Masks with Ties Level 3
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action The Therapeutic Goods Administration (TGA) conducted a post-market review of face masks included in the Australian Register of Therapeutic Goods (ARTG). Med-Con Pty Ltd was found to have supplied four batches of…
Check affected High Filtration Surgical Face Masks with Ties Level 3 identifiers →Therapeutic Goods Administration · July 29, 2026
- Affected product
- Rua Puawai THC24
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action Rua Puawai THC24 was intended for supply in the New Zealand market; however, affected stock was supplied in error to two Australian pharmacies in place of Rua Puawai T24. No other batches of Rua Puawai THC24 have…
Check affected Rua Puawai THC24 identifiers →Therapeutic Goods Administration · July 29, 2026
- Affected product
- SACRO Sacrocolpopexy Tips
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action Certain LOTs were assembled with an internal hub in the incorrect orientation. The inverted assembly prevents the tip from being attached to the compatible RUMI system or Advincula Arch handle. If the SACRO…
Check affected SACRO Sacrocolpopexy Tips identifiers →Therapeutic Goods Administration · July 28, 2026
- Affected product
- GORE EXCLUDER Thoracoabdominal Branch Endoprosthesis (TAMBE)
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Following a request from the US FDA to update the Instructions for Use (IFU) for the GORE EXCLUDER Thoracoabdominal Branch Endoprosthesis (TAMBE) during its review of the TAMBE 2026 IDE Annual Report (G150071/R061)…
Check affected GORE EXCLUDER Thoracoabdominal Branch Endoprosthesis (TAMBE) identifiers →Therapeutic Goods Administration · July 27, 2026
- Affected product
- Fludeoxyglucose (18F) Injection (FDG)
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Cyclotek was notified of a defective product (cracked vial), that was supplied to a customer site on 27 July 2026. The defect was identified upon receipt and was returned to the manufacturing facility. A new vial…
Check affected Fludeoxyglucose (18F) Injection (FDG) identifiers →Therapeutic Goods Administration · July 26, 2026
- Affected product
- Bendamustine HCL 159mg in 0.9% Sodium Chloride in Polyolefin Bag (Freeflex) for
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Upon receipt of compounded product, Bendamustine, the customer has identified a leaking product within the protective inner clear bag. The spill was contained within the bag and deemed not safe for patient use. The…
Check affected Bendamustine HCL 159mg in 0.9% Sodium Chloride in Polyolefin Bag (Freeflex) for identifiers →Therapeutic Goods Administration · July 3, 2026
- Affected product
- Panocell 16. An in vitro diagnostic medical device (IVD)
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action Antibody identification testing performed with Donor B11752 (cell 11 of lot 17453/17453Z) may show positive reactions in the presence of Anti-P1, while the negative antigen from the master list would suggest a…
Check affected Panocell 16. An in vitro diagnostic medical device (IVD) identifiers →Therapeutic Goods Administration · June 5, 2026
- Affected product
- Stella 2.0 Implant Orientation Diagram (IOD) Incorrect Axis
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action STAAR Surgical has become aware that a limited number of IODs generated have the risk of an IOD with an incorrect axis value that could be clinically significant. Specifically, per the review of data in Stella 2.0,…
Check affected Stella 2.0 Implant Orientation Diagram (IOD) Incorrect Axis identifiers →Therapeutic Goods Administration · April 29, 2026
- Affected product
- Fertilisation Medium FEM-50
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action The endotoxin testing method has changed due to the end of life of the testing kits used by Genea Biomedx for quantifying endotoxin levels. The Instructions for Use (IFU) was not updated to remove the mention of…
Check affected Fertilisation Medium FEM-50 identifiers →Therapeutic Goods Administration · November 4, 2025
- Affected product
- Getinge Assured Batch Monitoring PCD
- Category
- Medical device recalls
- Hazard
- Classification Class III Reason for the Action Getinge has identified a problem with certain Getinge Assured Batch Monitoring PCD units, where the required O-ring was not installed during manufacture following a customer complaint. The root cause was…
Check affected Getinge Assured Batch Monitoring PCD identifiers →Therapeutic Goods Administration · July 26, 2024
- Affected product
- McGRATH MAC video laryngoscope
- Category
- Medical device recalls
- Hazard
- Classification Class I Reason for the Action Medtronic have reported that the battery management system within the devices being removed (McGRATH MAC EMS Video Laryngoscope device, catalogue number 300-200-000; supplied under Stryker Catalogue number…
Check affected McGRATH MAC video laryngoscope identifiers →Therapeutic Goods Administration · September 4, 2019
- Affected product
- Steam Sterilizers with vertical moving doors
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Getinge Sterilization AB received a report of a user injury involving the autoclave door. The sterilisation technician activated the door lock while a basket was obstructing the door’s movement. Rather than…
Check affected Steam Sterilizers with vertical moving doors identifiers →