OFFICIAL-SOURCE RECALL · AUSTRALIA
GS1-HT Extraction Script for use within EasyScreen Sample Processing Recall: Classification Class II Reason
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- GS1-HT Extraction Script for use within EasyScreen Sample Processing Kit Workflow. An in vitro diagnostic medical device (IVD).
- Manufacturer
- See official notice
- Lot number
- numbers
- Batch number
- numbers
- Serial number / range
- Numbers
- Date code
- dates
- Expiration date
- dates
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Classification Class II Reason for the Action The impacted extraction scripts contain a missing section of error-handling code for sample positions 81–95. If a low-volume or missing-sample condition occurs in positions 81 - 95, the script may apply recovery logic intended for positions 1 - 80. This can cause the liquid handler to move to preceding specimen positions and aspirate material from an incorrect tube into the corresponding deep-well plate. To date, no adverse events have been reported due to this problem. Instructions for Customers/Users Customers are to: • Follow the interim measures outlined in the Customer Letter for continued use of the product prior to installation of the correction. • Genetic Signatures has amended the extraction-script error-handling logic for sample positions 81–95. • Authorised personnel will install the corrected script and verify and record the installed version before the systems are returned to routine use. • No EasyScreen Sample Processing Kit reagents or consumables need to be quarantined or returned because the physical kit components are not affected. Action Commencement Date 3/08/2026 Responsible Entity Genetic Signatures Ltd
Hazard
Classification Class II Reason for the Action The impacted extraction scripts contain a missing section of error-handling code for sample positions 81–95. If a low-volume or missing-sample condition occurs in positions 81 - 95, the script may apply recovery logic intended for positions 1 - 80. This can cause the liquid handler to move to preceding specimen positions and aspirate material from an incorrect tube into the corresponding deep-well plate. To date, no adverse events have been reported due to this problem. Instructions for Customers/Users Customers are to: • Follow the interim measures outlined in the Customer Letter for continued use of the product prior to installation of the correction. • Genetic Signatures has amended the extraction-script error-handling logic for sample positions 81–95. • Authorised personnel will install the corrected script and verify and record the installed version before the systems are returned to routine use. • No EasyScreen Sample Processing Kit reagents or consumables need to be quarantined or returned because the physical kit components are not affected. Action Commencement Date 3/08/2026 Responsible Entity Genetic Signatures Ltd
Official recall notice
GS1-HT Extraction Script for use within EasyScreen Sample Processing Kit Workflow. An in vitro diagnostic medical device (IVD).
Agency: Therapeutic Goods Administration
Official publication date: August 3, 2026