Therapeutic Goods Administration · September 16, 2024
BIOFIRE FILMARRAY TORCH System Base. An in vitro diagnostic medical Recall: Classification Class II Reason
- Affected product
- BIOFIRE FILMARRAY TORCH System Base. An in vitro diagnostic medical
- Category
- Medical device recalls
- Hazard
- Classification Class II Reason for the Action Power cycling can lead to arcing inside of the power switch and may result in carbon build-up and/or deformation of the power switch case. Carbon build-up and/or deformation may result in an open circuit causing…