Beckman Coulter Chemistry Analyzer AU480. An in vitro diagnostic Recall: Classification Class II Reason for

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryAustraliaPublishedJuly 30, 2026Last updatedJuly 30, 2026Official source last checkedAugust 15, 2026 at 7:44 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
Beckman Coulter Chemistry Analyzer AU480. An in vitro diagnostic medical device (IVD)
Manufacturer
See official notice
Lot number
numbers
Batch number
numbers
Serial number / range
Numbers
Date code
dates
Expiration date
dates

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Classification Class II Reason for the Action Beckman Coulter has determined that CHEMISTRY ANALYZER AU480 with Windows 10 IoT Enterprise 2021 LTSC PCs listed above may output the message “System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This problem is caused by SQL Server processing associated with transaction log auto-growth and is more likely to occur when a new index is not created daily. This is because the alarm and operation logs are deleted when the index is switched, and deleting multiple days’ worth of logs at once increases system load. If no alternative measurement method is available, a delay in reporting patient results may lead to a delay in diagnosis and/or patient treatment. This problem does not cause incorrect test results. To date, no adverse events have been reported due to this problem. Instructions for Customers/Users Customers are to: - Create a new index once or more per day at an appropriate time. Refer to “5.1 Start the System” of the IFU to complete this

Hazard

Classification Class II Reason for the Action Beckman Coulter has determined that CHEMISTRY ANALYZER AU480 with Windows 10 IoT Enterprise 2021 LTSC PCs listed above may output the message “System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This problem is caused by SQL Server processing associated with transaction log auto-growth and is more likely to occur when a new index is not created daily. This is because the alarm and operation logs are deleted when the index is switched, and deleting multiple days’ worth of logs at once increases system load. If no alternative measurement method is available, a delay in reporting patient results may lead to a delay in diagnosis and/or patient treatment. This problem does not cause incorrect test results. To date, no adverse events have been reported due to this problem. Instructions for Customers/Users Customers are to: - Create a new index once or more per day at an appropriate time. Refer to “5.1 Start the System” of the IFU to complete this

Official recall notice

Beckman Coulter Chemistry Analyzer AU480. An in vitro diagnostic medical device (IVD)

Agency: Therapeutic Goods Administration

Official publication date: July 30, 2026

Read the official AU_TGA_DRAC recall notice →