High Filtration Surgical Face Masks with Ties Level 3 Recall: Classification Class II Reason for the Action

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryAustraliaPublishedJuly 29, 2026Last updatedJuly 29, 2026Official source last checkedAugust 14, 2026 at 7:44 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
High Filtration Surgical Face Masks with Ties Level 3
Manufacturer
See official notice
Lot number
numbers, Numbers
Batch number
numbers, Numbers
Date code
dates, Dates
Expiration date
dates, Dates

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Classification Class II Reason for the Action The Therapeutic Goods Administration (TGA) conducted a post-market review of face masks included in the Australian Register of Therapeutic Goods (ARTG). Med-Con Pty Ltd was found to have supplied four batches of their Mask, surgical, single use, which failed the TGA's fluid resistance tests at level 3. The fluid resistance test measures the mask's ability to protect against the penetration of blood and/or body fluids, as outlined in the Australian Standard: AS 4381:2015. Updated 29/07/2026: The affected batches will be recalled. Instructions for Customers/Users Customers are advised to quarantine the affected batches. TGA and Sponsor are still investigating, and further information will be provided once the investigation is complete. Updated 29/07/2026: Customers are to return the affected batches as outlined in the customer letter. Action Commencement Date 29/07/2026 Responsible Entity Med-Con Pty Ltd

Hazard

Classification Class II Reason for the Action The Therapeutic Goods Administration (TGA) conducted a post-market review of face masks included in the Australian Register of Therapeutic Goods (ARTG). Med-Con Pty Ltd was found to have supplied four batches of their Mask, surgical, single use, which failed the TGA's fluid resistance tests at level 3. The fluid resistance test measures the mask's ability to protect against the penetration of blood and/or body fluids, as outlined in the Australian Standard: AS 4381:2015. Updated 29/07/2026: The affected batches will be recalled. Instructions for Customers/Users Customers are advised to quarantine the affected batches. TGA and Sponsor are still investigating, and further information will be provided once the investigation is complete. Updated 29/07/2026: Customers are to return the affected batches as outlined in the customer letter. Action Commencement Date 29/07/2026 Responsible Entity Med-Con Pty Ltd

Official recall notice

High Filtration Surgical Face Masks with Ties Level 3

Agency: Therapeutic Goods Administration

Official publication date: July 29, 2026

Read the official AU_TGA_DRAC recall notice →