Agilent Technologies, Inc PD-L1 IHC 22C3 pharmDx Recall: Agilent identified missing hyphens in the IFU for
Agilent Technologies, Inc PD-L1 IHC 22C3 pharmDx recall in Canada: Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- PD-L1 IHC 22C3 pharmDx
- Manufacturer
- See official notice
- Model number
- GE006
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx (GE006), revision 02 (PD005650). Due to an inadvertent formatting change in section "9.2 tissue sections," tissue-section thickness was incorrectly listed as 45 μm instead of 4-5 μm, and slide-storage temperature was incorrectly listed as 28 °c instead of 2-8 °c. This issue is limited to the electronic IFU (eIFU) associated with the three lots.
Hazard
Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx (GE006), revision 02 (PD005650). Due to an inadvertent formatting change in section "9.2 tissue sections," tissue-section thickness was incorrectly listed as 45 μm instead of 4-5 μm, and slide-storage temperature was incorrectly listed as 28 °c instead of 2-8 °c. This issue is limited to the electronic IFU (eIFU) associated with the three lots.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
PD-L1 IHC 22C3 pharmDx
Agency: Health Canada
Official publication date: October 28, 2026