Sysmex Corporation Automated Blood Coagulation Analyzers
OFFICIAL-SOURCE RECALL · CANADA
Sysmex Corporation Automated Blood Coagulation Analyzers Recall: In October 2026, the legal manufacturer Sysmex
Sysmex Corporation Automated Blood Coagulation Analyzers recall in Canada: In October 2026, the legal manufacturer Sysmex notified Siemens. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- Sysmex Corporation
- Product
- Automated Blood Coagulation Analyzers
- Manufacturer
- See official notice
- Model number
- BN343315, CL971029, AX801645, CV339763
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems. Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers. The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer’s investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used.
Hazard
In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems. Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers. The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer’s investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Automated Blood Coagulation Analyzers
Agency: Health Canada
Official publication date: November 6, 2026