▣ ScanRecallsGet the app ↗

OFFICIAL-SOURCE RECALL · CANADA

Andreas Hettich Gmbh ROTOFIX 32A Centrifuge Recall: The lid lock on some ROTOFIX 32A Centrifuges may not

Andreas Hettich Gmbh ROTOFIX 32A Centrifuge recall in Canada: The lid lock on some ROTOFIX 32A Centrifuges may not always engage securely. In. Check models and the official notice.

CountryCanadaPublishedNovember 6, 2026Last updatedNovember 6, 2026Official source last checkedNovember 6, 2026 at 3:00 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
Andreas Hettich Gmbh
Product
ROTOFIX 32A Centrifuge
Manufacturer
See official notice
Model number
HET-120601

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

The lid lock on some ROTOFIX 32A Centrifuges may not always engage securely. In these instances, the handle bar/lid latch could be open during a centrifuge run.

Hazard

The lid lock on some ROTOFIX 32A Centrifuges may not always engage securely. In these instances, the handle bar/lid latch could be open during a centrifuge run.

What consumers should do

Contact the manufacturer if you require additional information.

Official recall notice

ROTOFIX 32A Centrifuge

Agency: Health Canada

Official publication date: November 6, 2026

Read the official CANADA_MEDICAL_DEVICES recall notice →