MAQUET CARDIOPULMONARY GMBH RotaFlow Console Recall: During internal validation of new replacement
MAQUET CARDIOPULMONARY GMBH RotaFlow Console recall in Canada: During internal validation of new replacement batteries, MCP identified an. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- RotaFlow Console
- Manufacturer
- See official notice
- Model number
- 70104.3292
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.
Hazard
During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
RotaFlow Console
Agency: Health Canada
Official publication date: October 29, 2026