▣ ScanRecallsGet the app ↗

OFFICIAL-SOURCE RECALL · CANADA

Datascope Corp CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Recall: Getinge has identified that some CARDIOSAVE® devices

Datascope Corp CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon recall in Canada: Getinge has identified that some CARDIOSAVE®. Check models and the official notice.

CountryCanadaPublishedNovember 19, 2026Last updatedNovember 19, 2026Official source last checkedNovember 24, 2026 at 4:15 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
Datascope Corp.
Product
CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps
Manufacturer
See official notice
Model number
0998-00-0800-53, 0998-00-0800-65

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.

Hazard

Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.

What consumers should do

Contact the manufacturer if you require additional information.

Official recall notice

CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps

Agency: Health Canada

Official publication date: November 19, 2026

Read the official CANADA_MEDICAL_DEVICES recall notice →