Excelsior Medical Llc Medline 0.9% Sodium Chloride Injection, USP 10ml in a 10ml
OFFICIAL-SOURCE RECALL · CANADA
Heraeus Medical Components, LLC VascuPuncture™ PICC High Performance Guidewire Recall: Hydrophilic coated guidewire may swell
Heraeus Medical Components, LLC VascuPuncture™ PICC High Performance Guidewire recall in Canada: Hydrophilic coated guidewire may swell. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- VascuPuncture™ PICC High Performance Guidewire
- Manufacturer
- See official notice
- Model number
- 18070280NW-05MHP, 18145280NW-05MHP
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Hydrophilic coated guidewire may swell inconsistently when exposed to saline, causing coating clumping and potential particle detachment. Released particles could enter the bloodstream and potentially cause embolic events. Users should immediately discontinue use of affected devices, quarantine inventory, and follow recall instructions for return.
Hazard
Hydrophilic coated guidewire may swell inconsistently when exposed to saline, causing coating clumping and potential particle detachment. Released particles could enter the bloodstream and potentially cause embolic events. Users should immediately discontinue use of affected devices, quarantine inventory, and follow recall instructions for return.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
VascuPuncture™ PICC High Performance Guidewire
Agency: Health Canada
Official publication date: November 19, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.
Keep an eye on the products at home.
Scan food and household product barcodes, save your items and get notified when a future recall may match.