Canada Biogenix Unauthorized drugs
OFFICIAL-SOURCE RECALL · CANADA
Datascope Corp CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Recall: Getinge has identified that some CARDIOSAVE® devices
Datascope Corp CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon recall in Canada: Getinge has identified that some CARDIOSAVE®. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- Datascope Corp.
- Product
- CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps
- Manufacturer
- See official notice
- Model number
- 0998-00-0800-53, 0998-00-0800-65
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.
Hazard
Getinge has identified that some CARDIOSAVE® devices may contain a power management board (PMB) and/or solenoid control board (SCB) with one or more capacitors installed with reversed polarity. These capacitors may degrade over time, potentially causing CARDIOSAVE® device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown. The device may continue to function as intended, since the component degradation is not an immediate effect. Continued operation in this condition may eventually lead to a circuit malfunction, affecting critical system functions as outlined in the factors contributing to risk. Functional testing of the PMBs and SCBs may not identify every affected assembly unless capacitor polarity is specifically inspected and verified.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
CARDIOSAVE® Hybrid And CARDIOSAVE® Rescue Intra-Aortic Balloon Pumps
Agency: Health Canada
Official publication date: November 19, 2026
Read the official CANADA_HEALTH recall notice →SCAN ONCE. KEEP WATCHING.
Keep an eye on the products at home.
Scan food and household product barcodes, save your items and get notified when a future recall may match.