Heraeus Medical Components, LLC VascuPuncture™ PICC High Performance Guidewire
OFFICIAL-SOURCE RECALL · CANADA
Arrow International Llc Subsidiary Of Spring-Wire Guide, Pressure Injectable PICC Recall: Teleflex has initiated a voluntary recall for
Arrow International Llc Subsidiary Of Spring-Wire Guide, Pressure Injectable PICC recall in Canada: Teleflex has initiated a voluntary recall. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Spring-Wire Guide, Pressure Injectable PICC Kit, Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire
- Manufacturer
- See official notice
- Model number
- PW-18080-HN, PW-18080-HS, PW-16130-HN, PR-35552-HPHNL, PR-35541-HPHNL, PR-35552-HPHNM, PR-35541-HPHNM, PR-35563-HPHNL, PR-45552-HPHNL, PR-45552-HPHNM, PR-45563-HPHNL, PR-45541-HPHNM, PR-45563-HPHNM, PR-45541-HPHNL
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Teleflex has initiated a voluntary recall for specific lots of the above-listed products because there is a potential quality issue affecting the supplied hydrophilic coated spring wire guide (SWG) components. Teleflex has confirmed the coating may swell inconsistently when hydrated with saline leading to clumping of the coating and potential for detachment.
Hazard
Teleflex has initiated a voluntary recall for specific lots of the above-listed products because there is a potential quality issue affecting the supplied hydrophilic coated spring wire guide (SWG) components. Teleflex has confirmed the coating may swell inconsistently when hydrated with saline leading to clumping of the coating and potential for detachment.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Spring-Wire Guide, Pressure Injectable PICC Kit, Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire
Agency: Health Canada
Official publication date: November 18, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.