▣ ScanRecallsGet the app ↗

VERIFIED OFFICIAL-SOURCE INDEX

Gentherm Medical, LLC Recalls in Canada

Browse official recall notices for Canada, then compare the product and every listed identifier with the item you own.

Last updated: October 18, 2026 at 10:11 PM UTC · Official sources are checked throughout the day.

Check your product by barcode, model or lot →

SCAN ONCE. KEEP WATCHING.

Keep an eye on the products at home.

Scan food and household product barcodes, save your items and get notified when a future recall may match.

Download on the App StoreGet it on Google Play

1 current and recent recall

Showing 1–1 of 1 notices · Page 1 of 1

Health Canada · August 28, 2026

Gentherm Medical, LLC Blanketrol® II 115V and Blanketrol® III 115V Recall: Gentherm Medical has received complaints of

Affected product
Gentherm Medical, LLC Blanketrol® II 115V and Blanketrol® III 115V
Category
Product recall
Hazard
Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the…
Check affected Gentherm Medical, LLC Blanketrol® II 115V and Blanketrol® III 115V identifiers →

Browse product recalls by year

Download on the App StoreGet it on Google Play