Blanketrol® II 115V and Blanketrol® III 115V Recall: Gentherm Medical has received complaints of
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- Blanketrol® II 115V and Blanketrol® III 115V
- Manufacturer
- See official notice
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.
Hazard
Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.
Official recall notice
Blanketrol® II 115V and Blanketrol® III 115V
Agency: Health Canada
Official publication date: August 28, 2026
Read the official CANADA_HEALTH recall notice →SCAN ONCE. KEEP WATCHING.