Angiodynamics, Inc. Also D.B.A. As Various Clef des Champs Inc. natural health
OFFICIAL-SOURCE RECALL · CANADA
Quality In Flow(QinFlow) Ltd QinFlow Warrior Lite Base Unit Recall: The warrior lite base unit operates in conjunction
Quality In Flow(QinFlow) Ltd QinFlow Warrior Lite Base Unit recall in Canada: The warrior lite base unit operates in conjunction with a. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- QinFlow Warrior Lite Base Unit
- Manufacturer
- See official notice
- Model number
- QIF-03
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
The warrior lite base unit operates in conjunction with a battery and a disposable unit to warm blood and IV fluids prior to administration to patients. Following a technical root-cause investigation into field complaints involving thermal deformation of disposable units, quality in flow determined that an internal mechanical shield was incorrectly assembled on the Printed Circuit Board Assembly (PCBA) during manufacturing. Under harsh environmental conditions or significant mechanical impact, an incorrectly assembled shield may detach from the PCBA and cause an electrical short circuit.
Hazard
The warrior lite base unit operates in conjunction with a battery and a disposable unit to warm blood and IV fluids prior to administration to patients. Following a technical root-cause investigation into field complaints involving thermal deformation of disposable units, quality in flow determined that an internal mechanical shield was incorrectly assembled on the Printed Circuit Board Assembly (PCBA) during manufacturing. Under harsh environmental conditions or significant mechanical impact, an incorrectly assembled shield may detach from the PCBA and cause an electrical short circuit.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
QinFlow Warrior Lite Base Unit
Agency: Health Canada
Official publication date: November 24, 2026
Read the official CANADA_HEALTH recall notice →SCAN ONCE. KEEP WATCHING.
Keep an eye on the products at home.
Scan food and household product barcodes, save your items and get notified when a future recall may match.