Merit Medical Systems Inc Surfacer® Inside-Out® Access Catheter System Recall: Merit Medical Systems, Inc. (Merit) is conducting a
Merit Medical Systems Inc Surfacer® Inside-Out® Access Catheter System recall in Canada: Merit Medical Systems, Inc. (Merit) is conducting a. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Surfacer® Inside-Out® Access Catheter System
- Manufacturer
- See official notice
- Model number
- 600200/C
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Merit Medical Systems, Inc. (Merit) is conducting a voluntary recall of the Workstation Sheath and Dilator due to the radiopaque tip of the sheath detaching. To date, 4 (four) product complaints have been reported; 2 (two) of these product complaints were adverse events.
Hazard
Merit Medical Systems, Inc. (Merit) is conducting a voluntary recall of the Workstation Sheath and Dilator due to the radiopaque tip of the sheath detaching. To date, 4 (four) product complaints have been reported; 2 (two) of these product complaints were adverse events.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Surfacer® Inside-Out® Access Catheter System
Agency: Health Canada
Official publication date: November 10, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.