Health Canada · October 15, 2026
ROCHE DIAGNOSTICS GMBH cobas® c 303 analytical unit Recall: An issue has been identified where vacuum tubing of
- Affected product
- ROCHE DIAGNOSTICS GMBH cobas® c 303 analytical unit
- Category
- Medical device recalls
- Hazard
- An issue has been identified where vacuum tubing of the cell rinse unit of the cobas c 303 analytical unit prematurely deteriorated. Investigations confirmed the issue. In the event of small pinholes or cracks, no impact on analytical performance was…