OFFICIAL-SOURCE RECALL · CANADA
Integra Lifesciences Corporation Omni-Tract Surgical Retractor System Recall: Integra Lifesciences is issuing a correction to the
Integra Lifesciences Corporation Omni-Tract Surgical Retractor System recall in Canada: Integra Lifesciences is issuing a correction to the. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Omni-Tract Surgical Retractor System
- Manufacturer
- See official notice
- Model number
- Not applicable
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Integra Lifesciences is issuing a correction to the Integra® Omni-Tract® Surgical Retractor System - existing Instructions For Use (IFU) related to prevacuum sterilization dry time requirements, to ensure the Omni-Tract® devices and components are properly dried during sterilization to qualify current ISO standards. The current Omni-Tract® Surgical Retractor System IFU states a dry time of 30 minutes. The validated drying times utilizing standard pre-vacuum temperature and time standards were significantly longer and needed to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used. Update: Integra LifeSciences is issuing this additional updated correction to the Integra® Omni-Tract® Surgical Retractor System - Instructions for Use (IFU) adding prevacuum sterilization parameters when using a rigid container. Previously, in our June 2026 communication, Integra LifeSciences provided drying times solely for polypropylene wrapped sterilization trays informing you of three (3) the dry times (90 minutes, 110 minutes, or 120 minutes) depending on the sterilization cycle exposure time. This updated notification contains additional sterilization testing validated using a rigid container with the updated prevacuum sterilization cycle parameters.
Hazard
Integra Lifesciences is issuing a correction to the Integra® Omni-Tract® Surgical Retractor System - existing Instructions For Use (IFU) related to prevacuum sterilization dry time requirements, to ensure the Omni-Tract® devices and components are properly dried during sterilization to qualify current ISO standards. The current Omni-Tract® Surgical Retractor System IFU states a dry time of 30 minutes. The validated drying times utilizing standard pre-vacuum temperature and time standards were significantly longer and needed to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used. Update: Integra LifeSciences is issuing this additional updated correction to the Integra® Omni-Tract® Surgical Retractor System - Instructions for Use (IFU) adding prevacuum sterilization parameters when using a rigid container. Previously, in our June 2026 communication, Integra LifeSciences provided drying times solely for polypropylene wrapped sterilization trays informing you of three (3) the dry times (90 minutes, 110 minutes, or 120 minutes) depending on the sterilization cycle exposure time. This updated notification contains additional sterilization testing validated using a rigid container with the updated prevacuum sterilization cycle parameters.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Omni-Tract Surgical Retractor System
Agency: Health Canada
Official publication date: August 11, 2026