Omni-Tract Surgical Retractor System Recall: Integra Lifesciences is issuing a correction to the

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryCanadaPublishedAugust 11, 2026Last updatedAugust 11, 2026Official source last checkedAugust 12, 2026 at 2:37 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
Omni-Tract Surgical Retractor System
Manufacturer
See official notice
Serial number / range
number

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Integra Lifesciences is issuing a correction to the Integra® Omni-Tract® Surgical Retractor System - existing Instructions For Use (IFU) related to prevacuum sterilization dry time requirements, to ensure the Omni-Tract® devices and components are properly dried during sterilization to qualify current ISO standards. The current Omni-Tract® Surgical Retractor System IFU states a dry time of 30 minutes. The validated drying times utilizing standard pre-vacuum temperature and time standards were significantly longer and needed to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.Update: Integra LifeSciences is issuing this additional updated correction to the Integra® Omni-Tract® Surgical Retractor System - Instructions for Use (IFU) adding prevacuum sterilization parameters when using a rigid container. Previously, in our June 2026 communication, Integra LifeSciences provided drying times solely for polypropylene wrapped sterilization trays informing you of three (3) the dry times (90 minutes, 110 minutes, or 120 minutes) depending on the sterilization cycle exposure time. This updated notification contains additional sterilization testing validated using a rigid container with the updated prevacuum sterilization cycle parameters.

Hazard

Integra Lifesciences is issuing a correction to the Integra® Omni-Tract® Surgical Retractor System - existing Instructions For Use (IFU) related to prevacuum sterilization dry time requirements, to ensure the Omni-Tract® devices and components are properly dried during sterilization to qualify current ISO standards. The current Omni-Tract® Surgical Retractor System IFU states a dry time of 30 minutes. The validated drying times utilizing standard pre-vacuum temperature and time standards were significantly longer and needed to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.Update: Integra LifeSciences is issuing this additional updated correction to the Integra® Omni-Tract® Surgical Retractor System - Instructions for Use (IFU) adding prevacuum sterilization parameters when using a rigid container. Previously, in our June 2026 communication, Integra LifeSciences provided drying times solely for polypropylene wrapped sterilization trays informing you of three (3) the dry times (90 minutes, 110 minutes, or 120 minutes) depending on the sterilization cycle exposure time. This updated notification contains additional sterilization testing validated using a rigid container with the updated prevacuum sterilization cycle parameters.

Official recall notice

Omni-Tract Surgical Retractor System

Agency: Health Canada

Official publication date: August 11, 2026

Read the official CANADA_HEALTH recall notice →

Scan products you own

Scan products you own once and let RecallGuard monitor future official recalls.

Scan & Monitor Products

Browse connected recall information

Related recalls