▣ ScanRecallsGet the app ↗

OFFICIAL-SOURCE RECALL · CANADA

Philips Ultrasound Llc Lumify Ultrasound System Recall: Potential issues with the Lumify Diagnostic

Philips Ultrasound Llc Lumify Ultrasound System recall in Canada: Potential issues with the Lumify Diagnostic Ultrasound System which may. Check models and the official notice.

CountryCanadaPublishedOctober 5, 2026Last updatedOctober 5, 2026Official source last checkedOctober 18, 2026 at 10:11 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
Philips Ultrasound Llc
Product
Lumify Ultrasound System
Manufacturer
See official notice
Model number
3.0, LUMIFY ULTRASOUND SYSTEM, 4.0, 795005, 795161, 5.1.3

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Potential issues with the Lumify Diagnostic Ultrasound System which may prevent or delay scanning functionality: Due to a software issue, the Lumify application may not recognize a previously registered transducer and prompt for re-registration (outside of the defined criteria in the user guide) without warning prior to point of care use. Scanning functionality is prevented until the registration step has been completed; registration requires wi-fi or cellular signal access. Lumify devices used outside the hospital, such as in ambulances, aircraft, or field settings, are at greater risk due to reduced access to wi-fi/cellular. Philips has received reports that these erroneous re-registration events have prevented scanning in pre-hospital environments and delayed care to patients. These pre-hospital events occurred outside of wi-fi or cellular access. Additionally, Philips has identified that the connection between the Lumify transducer USB cable to a customer's tablet/ phone may disconnect or have an unstable connection during use which may disrupt continuous function of the device and result in a delay of care.

Hazard

Potential issues with the Lumify Diagnostic Ultrasound System which may prevent or delay scanning functionality: Due to a software issue, the Lumify application may not recognize a previously registered transducer and prompt for re-registration (outside of the defined criteria in the user guide) without warning prior to point of care use. Scanning functionality is prevented until the registration step has been completed; registration requires wi-fi or cellular signal access. Lumify devices used outside the hospital, such as in ambulances, aircraft, or field settings, are at greater risk due to reduced access to wi-fi/cellular. Philips has received reports that these erroneous re-registration events have prevented scanning in pre-hospital environments and delayed care to patients. These pre-hospital events occurred outside of wi-fi or cellular access. Additionally, Philips has identified that the connection between the Lumify transducer USB cable to a customer's tablet/ phone may disconnect or have an unstable connection during use which may disrupt continuous function of the device and result in a delay of care.

What consumers should do

Contact the manufacturer if you require additional information.

Official recall notice

Lumify Ultrasound System

Agency: Health Canada

Official publication date: October 5, 2026

Read the official CANADA_HEALTH recall notice →

SCAN ONCE. KEEP WATCHING.

Keep an eye on the products at home.

Scan food and household product barcodes, save your items and get notified when a future recall may match.

Download on the App StoreGet it on Google Play

Browse connected recall information

Download on the App StoreGet it on Google Play