Innovacon, Inc hCG One Step Pregnancy Tests Recall: Product may detect low levels of HCG below the
Innovacon, Inc hCG One Step Pregnancy Tests recall in Canada: Product may detect low levels of HCG below the stated assay sensitivity which. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- Innovacon, Inc.
- Product
- hCG One Step Pregnancy Tests
- Manufacturer
- See official notice
- Model number
- FHC-202, FHC-101
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Product may detect low levels of HCG below the stated assay sensitivity which may result in false positive test results in samples from non-pregnant individuals. The package insert is being updated to note that low HCG levels can occur at or after onset of perimenopause. Customers should consider this when interpreting results.
Hazard
Product may detect low levels of HCG below the stated assay sensitivity which may result in false positive test results in samples from non-pregnant individuals. The package insert is being updated to note that low HCG levels can occur at or after onset of perimenopause. Customers should consider this when interpreting results.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
hCG One Step Pregnancy Tests
Agency: Health Canada
Official publication date: October 5, 2026