FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version Recall: Under specific conditions

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryCanadaPublishedOctober 8, 2026Last updatedOctober 8, 2026Official source last checkedOctober 8, 2026 at 7:40 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3
Manufacturer
See official notice

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.

Hazard

Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.

Official recall notice

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3

Agency: Health Canada

Official publication date: October 8, 2026

Read the official CANADA_MEDICAL_DEVICES recall notice →