OFFICIAL-SOURCE RECALL · CANADA
Oxynov, Inc FreeO2 Automated Oxygen Therapy Device Adult & Pediatric Recall: A battery reading error may occur when certain
Oxynov, Inc FreeO2 Automated Oxygen Therapy Device Adult & Pediatric recall in Canada: A battery reading error may occur when certain Freeo2. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- Oxynov, Inc.
- Product
- FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
- Manufacturer
- See official notice
- Model number
- FO2-110-00
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
Hazard
A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Agency: Health Canada
Official publication date: October 6, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.
Keep an eye on the products at home.
Scan food and household product barcodes, save your items and get notified when a future recall may match.
