FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
OFFICIAL-SOURCE RECALL · CANADA
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric Recall: A battery reading error may occur when
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
- Manufacturer
- See official notice
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
Hazard
A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
Official recall notice
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Agency: Health Canada
Official publication date: October 6, 2026