FreeO2 Automated Oxygen Therapy Device Adult & Pediatric Recall: A battery reading error may occur when

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryCanadaPublishedOctober 6, 2026Last updatedOctober 6, 2026Official source last checkedOctober 6, 2026 at 3:22 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Manufacturer
See official notice

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.

Hazard

A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.

Official recall notice

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric

Agency: Health Canada

Official publication date: October 6, 2026

Read the official CANADA_MEDICAL_DEVICES recall notice →