ForeSight™ Oximeter Cable Recall: This recall is regarding inaccurate StO2 readings

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryCanadaPublishedAugust 21, 2026Last updatedAugust 21, 2026Official source last checkedAugust 21, 2026 at 2:44 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
ForeSight™ Oximeter Cable
Manufacturer
See official notice
Serial number / range
number

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

This recall is regarding inaccurate StO2 readings from the HEMFSM10 ForeSight™ Oximeter Cable due to mispositioning of the sensor cables. Edwards has identified that during manufacturing, the laterality identification labels may be missing or inaccurately placed on the device. There is potential that the labels on the green cables may have been switched, with the channel 1 cable labeled as channel 2, and the channel 2 cable labeled as channel 1. This will lead to a mismatch as to the left and right values displayed on the monitor and what the labels indicate on the cables. Should this potential issue occur during use, the values displayed on the monitor will be switched and will represent the opposite region being monitored. Therefore, there is a potential for incorrect diagnosis, leading to potential delay of appropriate treatment. If either of the labels are missing, or the flag label and/or the sensor label do not match the channel label on the housing, contact the Technical Support. If the flag and sensor labels correctly correspond to the channel labels, there is no further action required.

Hazard

This recall is regarding inaccurate StO2 readings from the HEMFSM10 ForeSight™ Oximeter Cable due to mispositioning of the sensor cables. Edwards has identified that during manufacturing, the laterality identification labels may be missing or inaccurately placed on the device. There is potential that the labels on the green cables may have been switched, with the channel 1 cable labeled as channel 2, and the channel 2 cable labeled as channel 1. This will lead to a mismatch as to the left and right values displayed on the monitor and what the labels indicate on the cables. Should this potential issue occur during use, the values displayed on the monitor will be switched and will represent the opposite region being monitored. Therefore, there is a potential for incorrect diagnosis, leading to potential delay of appropriate treatment. If either of the labels are missing, or the flag label and/or the sensor label do not match the channel label on the housing, contact the Technical Support. If the flag and sensor labels correctly correspond to the channel labels, there is no further action required.

What consumers should do

Additional information Details Original

Official recall notice

ForeSight™ Oximeter Cable

Agency: Health Canada

Official publication date: August 21, 2026

Read the official CANADA_HEALTH recall notice →