ForeSight™ Oximeter Cable Recall: This recall is regarding inaccurate StO2 readings
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- ForeSight™ Oximeter Cable
- Manufacturer
- See official notice
- Serial number / range
- number
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
This recall is regarding inaccurate StO2 readings from the HEMFSM10 ForeSight™ Oximeter Cable due to mispositioning of the sensor cables. Edwards has identified that during manufacturing, the laterality identification labels may be missing or inaccurately placed on the device. There is potential that the labels on the green cables may have been switched, with the channel 1 cable labeled as channel 2, and the channel 2 cable labeled as channel 1. This will lead to a mismatch as to the left and right values displayed on the monitor and what the labels indicate on the cables. Should this potential issue occur during use, the values displayed on the monitor will be switched and will represent the opposite region being monitored. Therefore, there is a potential for incorrect diagnosis, leading to potential delay of appropriate treatment. If either of the labels are missing, or the flag label and/or the sensor label do not match the channel label on the housing, contact the Technical Support. If the flag and sensor labels correctly correspond to the channel labels, there is no further action required.
Hazard
This recall is regarding inaccurate StO2 readings from the HEMFSM10 ForeSight™ Oximeter Cable due to mispositioning of the sensor cables. Edwards has identified that during manufacturing, the laterality identification labels may be missing or inaccurately placed on the device. There is potential that the labels on the green cables may have been switched, with the channel 1 cable labeled as channel 2, and the channel 2 cable labeled as channel 1. This will lead to a mismatch as to the left and right values displayed on the monitor and what the labels indicate on the cables. Should this potential issue occur during use, the values displayed on the monitor will be switched and will represent the opposite region being monitored. Therefore, there is a potential for incorrect diagnosis, leading to potential delay of appropriate treatment. If either of the labels are missing, or the flag label and/or the sensor label do not match the channel label on the housing, contact the Technical Support. If the flag and sensor labels correctly correspond to the channel labels, there is no further action required.
What consumers should do
Additional information Details Original
Official recall notice
ForeSight™ Oximeter Cable
Agency: Health Canada
Official publication date: August 21, 2026
Read the official CANADA_HEALTH recall notice →Scan products you own
Scan products you own once and let RecallGuard monitor future official recalls.
Scan & Monitor Products