OFFICIAL-SOURCE RECALL · CANADA
Biosphere Medical S.A Embosphere® Microspheres-Syringe Recall: Merit Medical Systems, Inc., on behalf of its
Biosphere Medical S.A Embosphere® Microspheres-Syringe recall in Canada: Merit Medical Systems, Inc., on behalf of its subsidiary, Biosphere. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Embosphere® Microspheres-Syringe
- Manufacturer
- See official notice
- Model number
- S82OGH
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Merit Medical Systems, Inc., on behalf of its subsidiary, Biosphere Medical, is voluntarily recalling specific lots of the Merit Embosphere® Microspheres due to a supplier-related defect that may affect the device packaging sterile barrier. Specifically, certain lots provided to Merit may contain a hole in the blister packaging. Merit has chosen to remove the units from the market and request that customers immediately stop using or distributing the affected lots and return the affected product to Merit. According to Merit's evaluation of health hazard(s), use of the affected product may result in exposure to pathogens/pyrogens, which could lead to infection. To date, Merit has not received any complaints or reports of patient harm or injury associated with this issue.
Hazard
Merit Medical Systems, Inc., on behalf of its subsidiary, Biosphere Medical, is voluntarily recalling specific lots of the Merit Embosphere® Microspheres due to a supplier-related defect that may affect the device packaging sterile barrier. Specifically, certain lots provided to Merit may contain a hole in the blister packaging. Merit has chosen to remove the units from the market and request that customers immediately stop using or distributing the affected lots and return the affected product to Merit. According to Merit's evaluation of health hazard(s), use of the affected product may result in exposure to pathogens/pyrogens, which could lead to infection. To date, Merit has not received any complaints or reports of patient harm or injury associated with this issue.
What consumers should do
Contact Merit Medical Canada.
Official recall notice
Embosphere® Microspheres-Syringe
Agency: Health Canada
Official publication date: August 19, 2026
Read the official CANADA_HEALTH recall notice →SCAN ONCE. KEEP WATCHING.