▣ ScanRecallsGet the app ↗

OFFICIAL-SOURCE RECALL · CANADA

Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Recall: Beckman Coulter identified a software

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryCanadaPublishedSeptember 10, 2026Last updatedSeptember 10, 2026Official source last checkedSeptember 10, 2026 at 4:41 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Analyzer
Manufacturer
See official notice

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.

Hazard

Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.

Official recall notice

Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Analyzer

Agency: Health Canada

Official publication date: September 10, 2026

Read the official CANADA_HEALTH recall notice →