BECKMAN COULTER, INC Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Recall: Beckman Coulter identified a software issue
BECKMAN COULTER, INC Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry recall in Canada: Beckman Coulter identified a software. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Analyzer
- Manufacturer
- See official notice
- Model number
- C63521, C63522, C63520, C63519
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.
Hazard
Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Analyzer
Agency: Health Canada
Official publication date: September 10, 2026