OFFICIAL-SOURCE RECALL · CANADA
Cynosure Lutronic Technology Corporation Cyanosure DermaV Laser System Recall: Cynosure Lutronic has identified a adverse trend of
Cynosure Lutronic Technology Corporation Cyanosure DermaV Laser System recall in Canada: Cynosure Lutronic has identified a adverse trend of. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- Cyanosure DermaV Laser System
- Manufacturer
- See official notice
- Model number
- 1011600210
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn.
Hazard
Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Cyanosure DermaV Laser System
Agency: Health Canada
Official publication date: August 26, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.