OFFICIAL-SOURCE RECALL · CANADA
Ge Medical Systems Information CareIntellect Perinatal Recall: GE Healthcare has become aware of an issue in
Ge Medical Systems Information CareIntellect Perinatal recall in Canada: GE Healthcare has become aware of an issue in CareIntellect. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Product
- CareIntellect Perinatal
- Manufacturer
- See official notice
- Model number
- K2504AA
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
GE Healthcare has become aware of an issue in CareIntellect Perinatal (CIP) that may result in loss of monitoring information and clearance of active patient monitoring alarms. When this issue occurs due to an interruption in system connectivity, monitoring data will be unavailable for up to approximately 20 seconds and currently active visual and audible alarm indicators will no longer be displayed or announced at the CIP user interface. Following resolution of system connectivity, alarm triggers will restart based on the configured thresholds. This issue could delay recognition of a maternal or fetal condition requiring clinical assessment or intervention.
Hazard
GE Healthcare has become aware of an issue in CareIntellect Perinatal (CIP) that may result in loss of monitoring information and clearance of active patient monitoring alarms. When this issue occurs due to an interruption in system connectivity, monitoring data will be unavailable for up to approximately 20 seconds and currently active visual and audible alarm indicators will no longer be displayed or announced at the CIP user interface. Following resolution of system connectivity, alarm triggers will restart based on the configured thresholds. This issue could delay recognition of a maternal or fetal condition requiring clinical assessment or intervention.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
CareIntellect Perinatal
Agency: Health Canada
Official publication date: October 1, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.