OFFICIAL-SOURCE RECALL · CANADA
Beckman Coulter AU Immunochemistry System - AU480 Analyzer Recall: Beckman Coulter has determined that a
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- Beckman Coulter AU Immunochemistry System - AU480 Analyzer
- Manufacturer
- See official notice
- Serial number / range
- number
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Beckman Coulter has determined that a software defect in AU480 chemistry analyzers using Microsoft® SQL server 2022 express may cause a DB connection error that renders alarmDB unreadable and prevents instrument startup. The issue is associated with transaction log auto-growth, and may cause delays in reporting patient test results to physicians.
Hazard
Beckman Coulter has determined that a software defect in AU480 chemistry analyzers using Microsoft® SQL server 2022 express may cause a DB connection error that renders alarmDB unreadable and prevents instrument startup. The issue is associated with transaction log auto-growth, and may cause delays in reporting patient test results to physicians.
Official recall notice
Beckman Coulter AU Immunochemistry System - AU480 Analyzer
Agency: Health Canada
Official publication date: August 6, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →Scan products you own
Scan products you own once and let RecallGuard monitor future official recalls.