Biomet Uk Limited Avantage Inserter Handles and Grippers Recall: It was identified while using the Avantage inserter
Biomet Uk Limited Avantage Inserter Handles and Grippers recall in Canada: It was identified while using the Avantage inserter handles and. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- Biomet Uk Limited
- Product
- Avantage Inserter Handles and Grippers
- Manufacturer
- See official notice
- Model number
- 110134058, 110134056, 110134044, 110134054, 110134052, 110134062, 110134064, 110134060, 110134046, 110134048, 110134050
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
It was identified while using the Avantage inserter handles and grippers during the implantation of the acetabular shell, that the acetabular shell may not consistently disengage from the gripper when the trigger is unlocked. As a result, the inserter handle and gripper assembly may remain engaged with the implanted shell, preventing clean separation of the instrumentation from the implant after its seating.
Hazard
It was identified while using the Avantage inserter handles and grippers during the implantation of the acetabular shell, that the acetabular shell may not consistently disengage from the gripper when the trigger is unlocked. As a result, the inserter handle and gripper assembly may remain engaged with the implanted shell, preventing clean separation of the instrumentation from the implant after its seating.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Avantage Inserter Handles and Grippers
Agency: Health Canada
Official publication date: October 8, 2026