Abiomed Inc Automated Impella® Controllers (AIC) Recall: Abiomed has decided to initiate a voluntary medical
Abiomed Inc Automated Impella® Controllers (AIC) recall in Canada: Abiomed has decided to initiate a voluntary medical device recall. Check models and the official notice.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- Abiomed Inc.
- Product
- Automated Impella® Controllers (AIC)
- Manufacturer
- See official notice
- Model number
- 0042-0040-CA, 0042-0000-CA, 0042-0000-US
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Abiomed has decided to initiate a voluntary medical device recall (removal) for design changes that were previously implemented through the Abiomed field service process. Abiomed performed a retrospective review of Field Change Notices (FCN) and determined that certain AIC consoles in the field did not receive the updates introduced by FCN 113. Failure to implement the applicable FCN 113 updates may result in the screen freezing during the Purge Cassette Change Wizard and the display of a false controller error. These defects could delay the initiation of support during device set-up and, in high-risk patient populations, may lead to temporary inadequate hemodynamic support or a temporary loss of hemodynamic support. Hospital inventory may continue to be used to ensure continuity of care. Not all units are affected by this issue.
Hazard
Abiomed has decided to initiate a voluntary medical device recall (removal) for design changes that were previously implemented through the Abiomed field service process. Abiomed performed a retrospective review of Field Change Notices (FCN) and determined that certain AIC consoles in the field did not receive the updates introduced by FCN 113. Failure to implement the applicable FCN 113 updates may result in the screen freezing during the Purge Cassette Change Wizard and the display of a false controller error. These defects could delay the initiation of support during device set-up and, in high-risk patient populations, may lead to temporary inadequate hemodynamic support or a temporary loss of hemodynamic support. Hospital inventory may continue to be used to ensure continuity of care. Not all units are affected by this issue.
What consumers should do
Contact the manufacturer if you require additional information.
Official recall notice
Automated Impella® Controllers (AIC)
Agency: Health Canada
Official publication date: October 2, 2026