OFFICIAL-SOURCE RECALL · CANADA
HeartSine® Samaritan® PAD 350P/360P/500P Recall: Stryker has identified a potential issue with the
HeartSine® Samaritan® PAD 350P/360P/500P recall in Canada: Stryker has identified a potential issue with the electrical connection. Read the official recall notice for details.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- HeartSine® Samaritan® PAD 350P/360P/500P
- Manufacturer
- See official notice
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Stryker has identified a potential issue with the electrical connection between Pad-Pak ™ or Pediatric-Pak™ and the AED. The connection uses spring-loaded electrical contacts, also known as pogo pins. Over time, oxidation of the pogo pin may increase electrical resistance and weaken the internal spring. If the condition progresses, the electrical connection between the Pad-Pak ™ or Pediatric-Pak™ battery and the AED may become intermittent or may be lost.
Hazard
Stryker has identified a potential issue with the electrical connection between Pad-Pak ™ or Pediatric-Pak™ and the AED. The connection uses spring-loaded electrical contacts, also known as pogo pins. Over time, oxidation of the pogo pin may increase electrical resistance and weaken the internal spring. If the condition progresses, the electrical connection between the Pad-Pak ™ or Pediatric-Pak™ battery and the AED may become intermittent or may be lost.
Official recall notice
HeartSine® Samaritan® PAD 350P/360P/500P
Agency: Health Canada
Official publication date: November 25, 2026
Read the official CANADA_MEDICAL_DEVICES recall notice →SCAN ONCE. KEEP WATCHING.