VELA Tango 700E Mobility Chair Recall: Classification Class II Reason for the Action In

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryAustraliaPublishedJanuary 1, 2036Last updatedJanuary 1, 2036Official source last checkedAugust 26, 2026 at 7:44 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
VELA Tango 700E Mobility Chair
Manufacturer
See official notice
Lot number
numbers
Batch number
numbers
Date code
dates, Date
Expiration date
dates, Date

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Classification Class II Reason for the Action In chairs manufactured between February 2025 and October 2025 the manufacturer, Vermund Larsen A/S, has identified a defect in the seat tilt actuator which may cause it to fracture during use. This may cause the seat to move unexpectedly, potentially resulting in the user falling or sustaining injury. Instructions for Customers/Users Customers should

Hazard

Classification Class II Reason for the Action In chairs manufactured between February 2025 and October 2025 the manufacturer, Vermund Larsen A/S, has identified a defect in the seat tilt actuator which may cause it to fracture during use. This may cause the seat to move unexpectedly, potentially resulting in the user falling or sustaining injury. Instructions for Customers/Users Customers should

Official recall notice

VELA Tango 700E Mobility Chair

Agency: Therapeutic Goods Administration

Official publication date: January 1, 2036

Read the official AU_TGA_DRAC recall notice →