PALACOS R+G pro 40, 60, 80 CE EUR and PALACOS MV+G pro 40, 60, 80 CE Recall: Classification Class II Reason
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- PALACOS R+G pro 40, 60, 80 CE EUR and PALACOS MV+G pro 40, 60, 80 CE EUR and COPAL G+C pro 40, 80 CE EUR
- Manufacturer
- See official notice
- Lot number
- numbers
- Batch number
- numbers
- Date code
- dates
- Expiration date
- dates
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Classification Class II Reason for the Action There have been isolated cases of monomer liquid leaking from the pro system. Assessments carried out to date have shown that the system can continue to be used as intended. However, if the product does leak, the system will need to be discarded and an alternate product used. If the product leaks, it cannot be entirely ruled out that users may come into
Hazard
Classification Class II Reason for the Action There have been isolated cases of monomer liquid leaking from the pro system. Assessments carried out to date have shown that the system can continue to be used as intended. However, if the product does leak, the system will need to be discarded and an alternate product used. If the product leaks, it cannot be entirely ruled out that users may come into
Official recall notice
PALACOS R+G pro 40, 60, 80 CE EUR and PALACOS MV+G pro 40, 60, 80 CE EUR and COPAL G+C pro 40, 80 CE EUR
Agency: Therapeutic Goods Administration
Official publication date: August 18, 2026