Orthopaedic implant components manufactured from ultra-high Recall: Classification Class II Reason for the

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryAustraliaPublishedAugust 11, 2026Last updatedAugust 11, 2026Official source last checkedAugust 21, 2026 at 7:46 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
Orthopaedic implant components manufactured from ultra-high molecular weight polyethylene (UHMWPE)
Manufacturer
See official notice
Lot number
numbers, Numbers
Batch number
numbers
Date code
dates
Expiration date
dates

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Classification Class II Reason for the Action Certain orthopaedic implants were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five (5) years. Industry guidance for storage duration of UHMWPE raw material used in the manufacture of implant bearing/articulating material of prosthetic joints is less than five (5) years. Product manufactured using UHMWPE raw material beyond the recommended storage duration has the potential for elevated levels of oxidation. Oxidation of UHMWPE material can impact material properties, potentially resulting in increased wear. Instructions for Customers/Users Upon receipt of the Recall letter, the customer is to: 1. Immediately review the Customer Letter and If

Hazard

Classification Class II Reason for the Action Certain orthopaedic implants were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five (5) years. Industry guidance for storage duration of UHMWPE raw material used in the manufacture of implant bearing/articulating material of prosthetic joints is less than five (5) years. Product manufactured using UHMWPE raw material beyond the recommended storage duration has the potential for elevated levels of oxidation. Oxidation of UHMWPE material can impact material properties, potentially resulting in increased wear. Instructions for Customers/Users Upon receipt of the Recall letter, the customer is to: 1. Immediately review the Customer Letter and If

Official recall notice

Orthopaedic implant components manufactured from ultra-high molecular weight polyethylene (UHMWPE)

Agency: Therapeutic Goods Administration

Official publication date: August 11, 2026

Read the official AU_TGA_DRAC recall notice →