Carescape Central Station (CSCS) V3.0.x Recall: Classification Class II Reason for the Action GE

Use the identifiers printed on your item to determine whether this recall may affect the product you own.

CountryAustraliaPublishedAugust 24, 2026Last updatedAugust 24, 2026Official source last checkedAugust 26, 2026 at 7:44 PM UTC

CHECK THE ITEM YOU OWN

Is your product affected?

Brand
See official notice
Product
Carescape Central Station (CSCS) V3.0.x
Manufacturer
See official notice
GTIN
00195278343512, 00195278343383
Lot number
numbers
Batch number
numbers
Date code
dates
Expiration date
dates

The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.

Why this product was recalled

Classification Class II Reason for the Action GE HealthCare has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the ‘Allow list database is corrupted’ message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events. There have been no injuries reported to GE HealthCare as a result of this problem. Instructions for Customers/Users - Customers are to review the actions required for continued use of the device in the Customer Letter. - GE HealthCare will provide a User Manual addendum that provides guidance regarding the prolonged startup condition. Action Commencement Date 24/08/2026 Responsible Entity GE Healthcare Australia Pty Ltd

Hazard

Classification Class II Reason for the Action GE HealthCare has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the ‘Allow list database is corrupted’ message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events. There have been no injuries reported to GE HealthCare as a result of this problem. Instructions for Customers/Users - Customers are to review the actions required for continued use of the device in the Customer Letter. - GE HealthCare will provide a User Manual addendum that provides guidance regarding the prolonged startup condition. Action Commencement Date 24/08/2026 Responsible Entity GE Healthcare Australia Pty Ltd

Official recall notice

Carescape Central Station (CSCS) V3.0.x

Agency: Therapeutic Goods Administration

Official publication date: August 24, 2026

Read the official AU_TGA_DRAC recall notice →