OFFICIAL-SOURCE RECALL · AUSTRALIA
Carescape Central Station (CSCS) V3.0.x Recall: Classification Class II Reason for the Action GE
Use the identifiers printed on your item to determine whether this recall may affect the product you own.
CHECK THE ITEM YOU OWN
Is your product affected?
- Brand
- See official notice
- Product
- Carescape Central Station (CSCS) V3.0.x
- Manufacturer
- See official notice
- GTIN
- 00195278343512, 00195278343383
- Lot number
- numbers
- Batch number
- numbers
- Date code
- dates
- Expiration date
- dates
The barcode may identify the product type, while the recall may apply only to certain lots, dates or variants. Check all identifiers listed in the official recall.
Why this product was recalled
Classification Class II Reason for the Action GE HealthCare has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the ‘Allow list database is corrupted’ message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events. There have been no injuries reported to GE HealthCare as a result of this problem. Instructions for Customers/Users - Customers are to review the actions required for continued use of the device in the Customer Letter. - GE HealthCare will provide a User Manual addendum that provides guidance regarding the prolonged startup condition. Action Commencement Date 24/08/2026 Responsible Entity GE Healthcare Australia Pty Ltd
Hazard
Classification Class II Reason for the Action GE HealthCare has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the ‘Allow list database is corrupted’ message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events. There have been no injuries reported to GE HealthCare as a result of this problem. Instructions for Customers/Users - Customers are to review the actions required for continued use of the device in the Customer Letter. - GE HealthCare will provide a User Manual addendum that provides guidance regarding the prolonged startup condition. Action Commencement Date 24/08/2026 Responsible Entity GE Healthcare Australia Pty Ltd
Official recall notice
Carescape Central Station (CSCS) V3.0.x
Agency: Therapeutic Goods Administration
Official publication date: August 24, 2026
Read the official AU_TGA_DRAC recall notice →Scan products you own
Scan products you own once and let RecallGuard monitor future official recalls.
Scan & Monitor Products